Description

  • Pulse 3a and Pulse 3b generating unit
  • A built-in 60V/30A Coupling Network allows independent operation.
  • <3.5ns rise time pulse available
  • Runs in pulse repetition frequencies sweep mode from 1 kHz to 100 kHz (ISS-7601 software required)

 

 

SPECIFICATION

 

Parameters

Specifications (Pulse 3a. 3b)

Output Voltage (Us)

Pulse 3a

- 20 ~  -350V ±10% ( -1V step)
 

Pulse 3b

20 ~  350V ±10%  (1V step)

Output Resistance (Ri)

50 ohm

Duration (td)

150ns±45ns

Rise time (tr)

Set as 5n

5ns±1.5ns (Us=±20V~±350V)

Set<3.5ns

< 3.5ns (Us=±50V~±300V)

Pulse period (t1)

10µs ~ 999µs ±10%  (1 µs step)

t4 (refer to the figures shown belown for its definition)

1ms ~ 100ms ±10% (1ms step)

t5(refer to the figures shown belown for its definition)

9ms ~ 999ms ±10% (1ms step)

Test Time

Set in min

1min ~ 999 min  ±10%  (1ms step)

Set in sec

1s ~ 999 s  ±10%  (1s step)

Circuit Breaker

Rated Current 40A

Interlock

External input:
Control at contact point or Open Collector
L Level: 0~0.4V       H level: 2V~4V

Remote control
from PC

Interface

Option: Optical Interface Unit Model: 07-00022A

Software

Option: Remote Control Software Model: ISS-7601

Compatible OS

Windows XP

DUT capacity

DC60Vmax, 30Amax

Power supply

AC100V, 110V, 120V, 200V, 220V, 240V ±10% (50/60Hz)

Operating temperature and humidity

 23°C±5°C / 25~75% (No dewing shall occur.)

Dimensions

(W) 430 X (H) 200 X (D) 522 mm
(Projections excluded)

Weight

Approx. 17kgs

Power consumption

Approx. 110VA

Standard accessories

Instruction manual
Replacement Fuse (3.16A)
Interlock plug
DC Input Cable (2m)
Output cable (0.5m)
BNC conversion adapter
50 ohm coaxial cable with BNC
GND cable
Short-circuit lead (for waveform verification

 

More products NOISEKEN

ESD ELECTROSTATIC DISCHARGE SIMULATOR NOISEKEN ESS-S3011

• 0.2 to 30.0kV +/-5% (30.5kV Maximum)
• 0.2 to 10kV: 0.05V step, 10.0 to 30.0kV: 0.1kV step
• EN/IEC 61000-4-2 Ed2 Compliant
• ISO 10605 Ed2 Compliant
• Pre-test
• Ten-key & Rotary knob to ease the setting
• Infra-red Remote Controller to allow remote Generator setting

 

ESD ELECTROSTATIC DISCHARGE SIMULATOR FOR SEMICONDUCTOR CIRCUITS NOISEKEN ESS-6002/ 6008

• ESS-6002 Output voltage: 10V ~2.0kV ±10% (1V step)
• ESS-6008 Output voltage: 100V ~ 8.0 kV±10% (10V step)
• A simple probe stand and test fixture plus a pair of miniature test leads
• Narrow pitch DUT can be tested in semi-automatic manner
• Each probe has a coaxial interface with an external measuring instrument.

 

PULSE 5A, 5B GENERATOR AUTOMOTIVE TRANSIENT SURGE SIMULATOR NOISEKEN ISS-7650

• Pulse 5a and Pulse 5b generating unit
• A built-in 60V/30A Coupling Network allows independent operation.
• Capacitive - resistive (inductive) discharge circuitry in strict conformity with the standard and with the actual car environment.
• Equipped with a programmable clip circuit that can produce Pulse 5b clipped waveform in steps of 0.1V without externally attaching a zener diode.
• Pulse 5b clipped waveform is pre-eminent in smoothness.


PULSE 1, 2A GENERATOR AUTOMOTIVE TRANSIENT SURGE SIMULATOR NOISEKEN ISS-7610

• Pulse 1 and Pulse 2a generating unit
• A built-in 60V/30A Coupling Network allows independent operation.
• Wide variety of selections of the output resistances Ri and Pulse Widths td and their combinations


RF EMISSION RADIATION PRECISION SCANNER NOISEKEN EPS-3007

• Provides an amplitude versus frequency plot for each antenna probe position and field intensity distribution map by a single scan
• Electric and magnetic fields are separately detected
• 3D graphical representation and zoom capability for a given area
• Comparison before and after a modification or improvement
• Scanning by a single antenna probe to minimize error between each measurement point
• 3GHz maximum frequency response (option)
• Standard included three kinds of antenna probes (Probe A, Probe B and Probe C)

 

ESD ELECTROSTATIC DISCHARGE SIMULATOR NOISEKEN ESS-B3011/ESS-L1611

• ESS-B3011 : 0.2 to 30.0kV
• ESS-L1611 : 0.2 to 16.0kV
• EN/IEC 61000-4-2 Ed2 Compliant (B3011/L1611)
• ISO 10605 Ed2 Compliant (B3011/L1611)
• 3 pre-checking function taking the confirmable test into the account (B3011/L1611)
• Discharge detecting function to realize the air-discharge confirmation (B3011/L1611)
• Lightest Discharge Gun in the market to lighten the continuous operation (GT-30R)

 

IMPULSE NOISE SIMULATOR EMC TESTING NOISEKEN INS-AX2

• INS-AX2-220/250 : 0.01 ~ 2.00kV±10% at 0.01kV step Pulse output voltage
• INS-AX2-420/450 : 0.01 ~ 4.00kV±10% at 0.01kV step Pulse output voltage
• Floating output
• 8 pulse widths installed: 10ns, 50ns, 100ns, 200ns, 400ns, 500ns, 800ns and 1000ns
• Internal switchable terminators
• Easily changeable and safety-interlocked outlet panel (EUT interface)
• 3 channels EUT fail input

 

IMPULSE NOISE SIMULATOR EMC TESTING NOISEKEN INS-4020/ 4040

• INS-4020 (2kV) : 0.01 ~ 2.00KV ±10% with 50 ohm load, 10V step
• INS-4040(4kV) : 0.01 ~ 4.00KV ±10% with 50 ohm load, 10V step
• Automatic ramp operation for output voltage, phase angle and repetition period
• Floating output
• Coupling mode selection by plugging the supplied coaxial connector to the selected port
• Easily changeable mercury relay unit

 

ESD ELECTROSTATIC DISCHARGE SIMULATOR NOISEKEN ESS-2000AX

• 0.20kV~30.0kV Output voltage range
• Fully programmable menu-driven simulator providing three operation modes: IEC standard, Manual and Sequence
• 5.7-inch color LCD for easy setting and good visibility
• Unique shape for the operator's easy access to the controls and displays even when the unit is put on the floor level (ground plane)
• Communications with PC through the optic serial interface
• wide variety of the dedicated options

 

LIGHTING SURGE SIMULATOR NOISEKEN LSS-6230

• Lightning Surge Simulator conforming to IEC61000-4-5 ed2 Standard
• Ring wave conforming to ANSI/IEEE C62.45 available
• Max. voltage 6.6kV output focusing on the practical use
• Monitor terminal standard equipped so as for easy waveform pre-checking
• Interlock built-in as the emergency stop function for safety