Description
SPECIFICATIONS
|
Parameters |
FNS-AX3-A16B |
FNS-AX3-B50B | |
|
Output voltage |
200 to 4800V | ||
|
Pulse parameters at 50 ohm |
Pulse rise time 5ns ±30%, Pulse width 50ns ±30% *1 | ||
|
Pulse parameters at 1k ohm |
Pulse rise time 5ns ±30%, Pulse width 35 ns to 150 ns *1 | ||
|
Polarity |
Positive, negative, polarity change burst by burst | ||
|
Pulse frequency |
0.1kHz to 2MHz *1 | ||
|
Burst duration |
1 to 1000 pulses or continuous 1 pulse per 1ms burst period | ||
|
Burst period |
10 to 1000 ms ±10%, 10ms step | ||
|
Continuous mode operation |
10 KHz max (0.2 to 1kV) | ||
|
Dynamic output impedance |
50 ohms | ||
|
Output connector |
Coaxial (PULSE OUT) | ||
|
Coupling/DC blocking capacitor |
33nF/10nF | ||
|
Frequency modulation |
0 to -10% of setting | ||
|
Timer (test duration) |
1s to 9h 59m 59s | ||
|
External trigger |
TL down edge Minimum trigger interval acceptable: 10ms | ||
|
Burst timing |
Non-synchronization |
Irrelevant to EUT mains frequency | |
|
Synchronization |
Synchronizes with the EUT mains frequency 0 to 360 degrees ±10 degrees, 1 degree step | ||
|
Internal CDN |
85 to 240VAC 16A 50/60Hz |
85 to 500VAC 50A 50/60Hz | |
|
Coupling line |
L, N, PE |
L1, L2, L3, N, PE | |
|
Coupling mode |
Common mode | ||
|
Dimensions (projection included) |
(W) 320 x (D)352 x (H)400 mm | ||
|
Character display |
English, Chinese, Korean or Japanese | ||
|
Weight |
Approx. 14 kg |
Approx. 19 k | |
*1 Pulse specifications guaranteed over 5KHz to 100KHz
Recommended options
More products NOISEKEN
|
PCB BOARDS RADIATION EMISSIONS MEASUREMENT SYSTEM NOISEKEN EPS-6000
• Measurable up to 8GHz in the frequency range |
|
VOLTAGE DIP AND SWELL WAVEFORMS SIMULATOR NOISEKEN VDS-2002
• Conforming to IEC 61000-4-11 |
|
ESD ELECTROSTATIC DISCHARGE SIMULATOR FOR SEMICONDUCTOR CIRCUITS NOISEKEN ESS-6002/ 6008
• ESS-6002 Output voltage: 10V ~2.0kV ±10% (1V step) |
|
ESD ELECTROSTATIC DISCHARGE SIMULATOR NOISEKEN ESS-B3011/ESS-L1611
• ESS-B3011 : 0.2 to 30.0kV |
|
RF EMISSION RADIATION PRECISION SCANNER NOISEKEN EPS-3007
• Provides an amplitude versus frequency plot for each antenna probe position and field intensity distribution map by a single scan |
|
IMPULSE NOISE SIMULATOR EMC TESTING NOISEKEN INS-4020/ 4040
• INS-4020 (2kV) : 0.01 ~ 2.00KV ±10% with 50 ohm load, 10V step |
|
HIGH VOLTAGE LIGHTING SURGE SIMULATOR NOISEKEN LSS-F02
• "Output voltage 15kV, current 7500A" which can conduct breakdown resistibility test |
|
FAST TRANSIENT BURST SIMULATOR NOISEKEN FNS-AX3
• 4.8kV maximum peak voltage |
|
PULSE 3A, 3B GENERATOR AUTOMOTIVE TRANSIENT SURGE SIMULATOR NOISEKEN ISS-7630
• Pulse 3a and Pulse 3b generating unit |
|
ESD ELECTROSTATIC DISCHARGE SIMULATOR NOISEKEN ESS-2000AX
• 0.20kV~30.0kV Output voltage range |