STANDARD RESISTORS

On request various values and precision described below and boxes containing several values at your choice

Included certificate of calibration of our laboratory

Fixed value standard resistances

high-accuracy used as standard in laboratory.

SAMAR certificate of calibration is supplied with each standard resistor.

The standard production includes sample resistances multiple of 10.

 

  • Resistances realized in manganine with temperature coefficient of ± 20 ppm/°C (independent from temperature between 15 and 30 °C)
  • Long term stability:  ± 50 ppm per year
  • Resistances from 0.0001? to 0.1? are 4 terminals, two potential and two current.
  • Resistances from 1? to 10M? are wrapped in an ant-inductive way and are two terminals with 2 screws for potential and current.

 

 

Type

Nominal Value

O

Max Current in Air,  A

Power dissipation in Air, W

Precision %

t=20 ± 0,5 °C

Max deviation from nominal value

t= 20 ± 0,5 °C

RC-E 50/00

0,0001

100

1

± 0,05

± 0,1

RC-E 50/0

0,001

30

1

± 0,01

± 0,05

RC-E 50/1

0,01

10

1

± 0,01

± 0,05

RC-E 50/2

0,1

3

1

± 0,01

± 0,05

RC-E 50/3

1

1

1

±0,01

± 0,05

RC-E 50/4

10

0,3

1

± 0,01

± 0,05

RC-E 50/5

100

0,1

1

± 0,01

± 0,05

RC-E 50/6

1.000

0,03

1

± 0,01

± 0,05

RC-E 50/7

10.000

0,01

1

± 0,01

± 0,05

RC-E 50/8

100.000

0,003

1

± 0,01

± 0,05

RC-E 50/9

1.000.000

0,001

1

± 0,01

± 0,05

RC-E 50/10

10.000.000

Vmax=500V

Vmax=500V

± 0,01

± 0,05

 

Alte produse SAMAR

CAPACITOR STANDARD DE VALOARE FIXA SAMAR CC-E50

• Values from 1nF to 1µF
• ON REQUEST DIFFERENT VALUES FROM THE ONES SHOWN
• Nominal voltage : 500Vdc
• Temperature coefficient: - (100 ± 50) x 10-6 /°C
• Test voltage: 2,000Vdc
• Insulation resistance: 500,000 MW
• Precision: ± 0.5 % at 1,000 Hz, between terminals

 

INDUCTANTA ETALON DE PRECIZIE PENTRU LABORATOR CALIBRARE SAMAR IC

• Values from 1mH to 1H
• Inductances of precision used as standard in laboratory
• SAMAR certificate of calibration is supplied with each standard inductor.
• Precision: ± 0.1 % at 1,000 Hz, between terminals
• ON REQUEST DIFFERENT VALUES FROM THE ONES SHOWN

 

DECADA REZISTENTE DE PRECIZIE SAMAR CR-E

• DECADES BOXES OF RESISTANCE with switch precision 0.1% at terminals
• Calibration precision: ± 0.02 %
• The precision shown refers to the terminals and therefore is total resistance included the resistance of zero
• Resistors are in manganine wire (temperature coefficient ± 10 ppm/°C) wrapped in ant-inductive way.
• Thanks to an opportune seasoning process the long term stability is ensured: : 10 ppm in the first year and max. 20 ppm in three years

 

REFERINTA STANDARD DE TENSIUNE FIXA DE PRECIZIE PENTRU CALIBRARE SAMAR SC51/SC52

• REFERENCES DC VOLTAGE STANDARD
• certificate of calibration of our laboratory
• DC voltage reference standard providing standard cell accuracy for use in measuring laboratories.
• Nominal e.m.f.: 1.01865 calibration 5°/ 6° decimal digits
• Precision: ± 50 µV / ± 25 µV
• max output current: 1 mA continuative
• stability: < 10 ppm/month with working temperature 15 - 25 °C and relative
• humidity 45-85%

 

CASETA REZISTENTE IN TREPTE DIVIZOR DE TENSIUNE SI POTENTIAL SAMAR CRC/1

• Used as a potential divider or for calibration (for example insulation tester)
• Chemical resistances of high stability, with dissipation 1W
• Precision 1% dc and ac up to 1 MHz. (on request execution at 0.1% )
• Range: 0 to 500M? with terminals for the following values:
• 0.5 - 1 - 2 - 3 - 4 - 5 - 10 - 20 -30 - 40 - 50 - 100 - 200 - 300 - 400 - 500 K?
• 1 - 2 - 3 - 4 - 5 - 10 - 20 -30 - 40 - 50 - 100 - 200 - 300 - 400 - 500 M?
• Execution in reading-desk metal case.

 

REZISTENTA ETALON STANDARD DE PRECIZIE UTILIZARE IN AER SAMAR RC-E 50

• On request various values and precision
• Long term stability: ± 50 ppm per year
• Resistances from 0.0001? to 0.1? are 4 terminals, two potential and two current.
• Resistances from 1? to 10M? are wrapped in an ant-inductive way and are two terminals with 2 screws for potential and current
• Included certificate of calibration of our laboratory
• high-accuracy used as standard in laboratory.

 

CAPACITATE ETALON DE PRECIZIE PENTRU LABORATOR CALIBRARE SAMAR CC

• Values from 1nF to 1µF
• ON REQUEST DIFFERENT VALUES FROM THE ONES SHOWN
• Nominal voltage: 500Vdc
• Temperature coefficient: - (100 ± 50) x 10-6 /°C
• Test voltage: 2,000Vdc
• Insulation resistance: 500,000 MW
• Precision ± 0.1 % at 1,000 Hz, between terminals

 

REZISTOR ETALON STANDARD REZISTENTE MARI SI INALTA TENSIUNE SAMAR RC80/M

• On request various values and precision
• Nominal Values: 10MO..100TO
• Nominal Voltage 500V
• Included certificate of calibration of our laboratory

 

INDUCTOR STANDARD PENTRU LABORATOARE CALIBRARE SAMAR IC-E50

• Values from 1mH to 1H
• Inductances of precision used as standard in laboratory
• SAMAR certificate of calibration is supplied with each standard inductor.
• Precision: ± 0.5 % at 1,000 Hz, between terminals
• ON REQUEST DIFFERENT VALUES FROM THE ONES SHOWN

 

PUNTE DE MASURA WHEATSTONE THOMSON MAXWELL DE SAUTY SCHERING

• DC / AC BRIDGES
• THOMSON - WHEATSTONE BRIDGE
• MAXWELL - DE SAUTY BRIDGE 50Hz
• SCHERING BRIDGE