STANDARD RESISTORS
On request various values and precision described below and boxes containing several values at your choice
Included certificate of calibration of our laboratory
Fixed value standard resistances
high-accuracy used as standard in laboratory.
SAMAR certificate of calibration is supplied with each standard resistor.
The standard production includes sample resistances multiple of 10.
-
Long term stability: +/- 10 ppm per year
-
Resistances from 0.0001? to 10? are 4 terminals
-
Resistances from 100? to 10M? are two electrolytic copper terminals with 2 screws, one for current and one for voltage
-
The potential terminals are screwed with plug connection.
-
Current terminals are of electrolytic copper widely sizeable.
|
Type |
Nominal Value
O |
Max Current in Air, A |
Max Current in Oil, A |
Power dissipation in Air, W |
Precision %
t= 20 ± 0,5 °C |
Max deviation from nominal value
t= 20 ± 0,5 °C |
|
RC80/1 |
0,0001 |
300 |
600 |
9 |
± 0,01 |
± 0,05 |
|
RC80/2 |
0,001 |
75 |
200 |
5,6 |
± 0,01 |
± 0,02 |
|
RC80/3 |
0,01 |
30 |
70 |
9 |
± 0,005 |
± 0,01 |
|
RC80/4 |
0,1 |
5 |
15 |
2,5 |
± 0,002 |
± 0,005 |
|
RC80/5 |
1 |
1,5 |
4 |
2,25 |
±0,002 |
± 0,005 |
|
RC80/6 |
10 |
0,5 |
2,5 |
2,5 |
± 0,002 |
± 0,005 |
|
RC80/7 |
100 |
0,2 |
0,5 |
4 |
± 0,002 |
± 0,005 |
|
RC80/8 |
1.000 |
0,07 |
0,15 |
5 |
± 0,002 |
± 0,005 |
|
RC80/9 |
10.000 |
0,015 |
0,03 |
2,25 |
± 0,002 |
± 0,005 |
|
RC80/10 |
100.000 |
0,005 |
0,01 |
2,5 |
± 0,005 |
± 0,01 |
|
RC80/11 |
1.000.000 |
0,0005 |
0,01 |
0,25 |
± 0,005 |
± 0,02 |
|
Alte produse SAMAR
DECADA CAPACITATI DE PRECIZIE SAMAR CC-B
• certificate of calibration of our laboratory • DECADES BOXES OF CAPACITY with switch precision 2% at terminals • Polypropylene capacitor • Tangd : 0.003 to 1KHz • Precision: to 1.000Hz at terminals ±2% • Working voltage: 400Vdc max; 100V switching
|
CAPACITOR STANDARD DE VALOARE FIXA SAMAR CC-E50
• Values from 1nF to 1µF • ON REQUEST DIFFERENT VALUES FROM THE ONES SHOWN • Nominal voltage : 500Vdc • Temperature coefficient: - (100 ± 50) x 10-6 /°C • Test voltage: 2,000Vdc • Insulation resistance: 500,000 MW • Precision: ± 0.5 % at 1,000 Hz, between terminals
|
BOXA DECADICA REZISTENTA DE PRECIZIE SAMAR CR
• DECADES BOXES OF RESISTANCE with switch precision 0.05% at terminals • Calibration precision: ± 0.02 % • Thanks to an opportune seasoning process the long term stability is ensured: 10 ppm in the first year and max 20 ppm in three years. • Switches have a resistance of contact lower than 0.001 W invariable in the time. • certificate of calibration of our laboratory
|
CASETA REZISTENTE IN TREPTE DIVIZOR DE TENSIUNE SI POTENTIAL SAMAR CRC/1
• Used as a potential divider or for calibration (for example insulation tester) • Chemical resistances of high stability, with dissipation 1W • Precision 1% dc and ac up to 1 MHz. (on request execution at 0.1% ) • Range: 0 to 500M? with terminals for the following values: • 0.5 - 1 - 2 - 3 - 4 - 5 - 10 - 20 -30 - 40 - 50 - 100 - 200 - 300 - 400 - 500 K? • 1 - 2 - 3 - 4 - 5 - 10 - 20 -30 - 40 - 50 - 100 - 200 - 300 - 400 - 500 M? • Execution in reading-desk metal case.
|
REFERINTA STANDARD DE TENSIUNE FIXA DE PRECIZIE PENTRU CALIBRARE SAMAR SC51/SC52
• REFERENCES DC VOLTAGE STANDARD • certificate of calibration of our laboratory • DC voltage reference standard providing standard cell accuracy for use in measuring laboratories. • Nominal e.m.f.: 1.01865 calibration 5°/ 6° decimal digits • Precision: ± 50 µV / ± 25 µV • max output current: 1 mA continuative • stability: < 10 ppm/month with working temperature 15 - 25 °C and relative • humidity 45-85%
|
REZISTOR DE PRECIZIE IN DECADE SAMAR CR-B
• Power: 0.5 W permanent • Resistances: 0.1W and 1 W to coil, from 10W to 1 MW metal film to 50ppm • Security rule terminals • Plastic box, panel of aluminium with serigraphy • Double insulation • Work frequency: 0 - 500 KHz • Precision: 0.5 %
|
INDUCTOR STANDARD PENTRU LABORATOARE CALIBRARE SAMAR IC-E50
• Values from 1mH to 1H • Inductances of precision used as standard in laboratory • SAMAR certificate of calibration is supplied with each standard inductor. • Precision: ± 0.5 % at 1,000 Hz, between terminals • ON REQUEST DIFFERENT VALUES FROM THE ONES SHOWN
|
|