  • Total resistance range 10.00 k? to 1000.0 G?
  • Resistance accuracy 0.1 to 5 %
  • Test voltage range 10 000 VDC + 5%
  • Test voltage accuracy indication 0.5 % + 10 V
  • Test current indication 4 digit from 0.01 pA to 99.99 mADC
  • Programmable high resistance decade box
  • Short current verification
  • Dielectric and absorption parameters simulation
  • High voltage capacitance in three fix points 10, 50, 100 nF
  • HIPOT timer verification 

 

 

Description

M191 calibrator is based on programmable high resistance decade in summary range from 10 k? to 1 T?. The decade is designed for operation voltages up to 10 kV. In this range it offers basic accuracy 0.1 % to 5 % depending on set resistance value. 
Device is completed with additional electronic circuits allowing calibration not only resistance ranges, but also calibration of test voltage generated by UUT, short current verifying, verifying of functions of measuring dielectric parameters like polarization index (PI), dielectric absorption ratio (DAR) and polarization ratio (PR). The calibrator enables also verification of timer function of UUTs. 

The calibration can be controlled manually using front panel keypad or in remote mode using one of two types of interfaces GPIB, RS232. The calibrator can easily fit within calibration systems featuring CALIBER software support.

 

 

Alte produse MEATEST

DECADA REZISTIVA CALIBRARE MEGOHMMETRE SI TESTERE IZOLATIE 5KV MEATEST M109R

• Calibration of insulation meters and megaohm meters
• Four decades
• Resistance range 1 MO - 12 GO
• Accuracy from 0.1 % to 1 %
• Maximum working voltage 5 kVdc
• Internal battery / power line adapter supplying
• RS232 bus

 

REZISTENTA CAPACITATE INDUCTANTA STANDARD PENTRU CALIBRARE RLC-METRE

• Resistance Capacitance Inductance standards
• Traceability of electric quantities
• Calibration of meters
• Calibration uncertainty 0.005 %
• Low time constant
• Three/four terminal or four pair terminal connection

 

DECADA CAPACITIVA CALIBRARE RLC-METRE MEATEST M520

• Designed for RCL meters calibration
• Range of capacitance from 100 pF to 12 mF
• Basic accuracy 0.25 %
• Maximal operating voltage 50 V
• Power supplying from internal battery or power line adapter
• Remote control RS-232

 

CALIBRATOR UNIVERSAL MULTIFUNCTIONAL DE LABORATOR SI PRECIZIE MEATEST M143

• Wide voltage range from 0 to 1000 V DC/AC with accuracy 0.01 %
• Extended current range from 0 to 20 A DC/AC with accuracy 0.02 %
• Extended current range to 1000 A for calibration of clamp ammeters
• Sinusoidal & Non-sinusoidal waveforms
• Fix standard resistors 10 O to 100 MO in decimal values with calibration accuracy 0.02 %
• Thermocouple temperature sensor simulation R, S, B, J, T, E, K, N in range from 250 °C to 1850 °C
• Cold junction automatic compensation using external Pt1000 sensor
• RTD temperature sensor simulator as option

 

CALIBRATOR DE PUTERE SI ENERGIE ELECTRICA MEATEST M133

• Voltage "phase-neutral" 1 V to 280 Vdc / 300 mA, 1 V to 600 Vac
• AC voltage "phase - phase" in the 3phase system 2 V to 1200 V
• DC/AC current 8 mA to 30 A / 5 V
• Power factor setting -1 to +1
• Harmonic, interharmonic distortion, modulation
• Frequency range 15 Hz to 1 kHz
• Simulated electric power to 8.4 kVA (420 kVA with option 140-50)
• Built-in process multimeter
• GPIB & RS232 interface

 

BOXA REZISTENTE IN DECADE PROGRAMABILA DE PRECIZIE MEATEST M630

• Resistance range 1.000 00 O - 1.200 000 MO
• Accuracy 0.003%
• Maximal voltage 200Vpk
• Resolution from 10 µO
• Range of Pt sensor temperature simulation -200°C - +850°C
• Connection 2, 3, 4-wire
• Remote control RS232, optionally USB, IEEE488, Ethernet

 

DECADA REZISTIVA 10KV CALIBRARE MEGOHMMETRE SI TESTERE IZOLATIE MEATEST M191

• Total resistance range 10.00 k? to 1000.0 G?
• Resistance accuracy 0.1 to 5 %
• Test voltage range 10 000 VDC + 5%
• Test voltage accuracy indication 0.5 % + 10 V
• Test current indication 4 digit from 0.01 pA to 99.99 mADC
• Programmable high resistance decade box
• Short current verification
• Dielectric and absorption parameters simulation: polarization index (PI), dielectric absorption ratio (DAR) and polarization ratio (PR)
• High voltage capacitance in three fix points 10, 50, 100 nF
• HIPOT timer verification

 

SARCINA REZISTIVA AC/DC CALIBRARE SURSE TENSIUNE, TRANSFORMATOARE MEATEST M192

• Resistance range from 15 O to 300 kO
• Maximal load 3kW, maximal voltage 250 Vrms
• Accuracy 0.1 %
• Temperature coefficient < 10 ppm/°C
• Interface RS - 232 (IEEE488 optionally)
• Table version (19" module as option)

 

DECADA REZISTIVA PROGRAMABILA MEATEST M602

• Resistance range from 0.1 O to 10 MO with accuracy 0.02 %
• Temperature coefficient < 25 ppm/°C
• Max. load 2 W (120 V and 0.6 A)
• RTD (Pt, Ni) simulation with accuracy 0.2 °C
• Power supplying from power line adaptor or from internal battery
• Interface RS232 (IEEE488 optionally)
• Table version (19" module as option)

 

CALIBRATOR PUTERE SI ENERGIE ACTIVA, REACTIVA, APARENTA, FAZA, UNGHI MEATEST M133I

• Voltage "phase-neutral" 1 V to 280 Vdc / 300 mA, 1 V to 600 Vac
• AC voltage "phase - phase" in the 3phase system 2 V to 1200 V
• DC/AC current 30 mA to 30 A / 5 V, accuracy 310 ppm
• Power factor setting -1 to +1
• Frequency range 15 Hz to 1 kHz
• Simulated electric power up to 8.4 kVA (420 kVA with option 140-50)
• Built-in process multimeter
• GPIB & RS232 interface