Description
Operating principle
Single-channel Gaussmeter with a resolution of 4 ¾ digits with filter in DC-adjustment and a measuring accuracy of ± 0.20 % of the measured value. The measuring rate is 3 measurements per second (display) and 30 measurements per second with USB interface. Measurement of Hall voltage created in Hall sensor which is proportional to the applied magnetic field.
All the functions are displayed and controlled via the front panel of the instrument. Input via keyboard, readings on a 2-line LCD display with background light. Two analogue outputs are provided. The corrected analogue output displays the exact value, the non-corrected analogue output shows the field strength in real time. Field and temperature measurements are via the probe input.
The Gaussmeter 425 DSP is used e.g. for testing magnetic materials in sensor technology and in automation. It can also function as an integrated component in controlling production processes and in the manufacture of DC motors.
|
Technical data: |
Gaussmeter 425 |
|
Measuring mode: |
measurement of Hall voltage |
|
Measuring range: |
350 mG, 3.5 G, 35 G, 350 G, 3.5 kG, 35 kG, 350 kG |
|
Measuring frequency: |
DC, AC up to 10 kHz |
|
Measuring accuracy: |
± 0,2 % of DC value, ± 2 % of AC value 0,005 % of measuring range |
|
Repeatability: |
± 0,1 % |
|
Measuring rate: |
3 measurements/sec / display |
|
Selection of measuring range: |
automatic or manual |
|
Resolution: |
4 3/4 digits |
|
Display: |
2-line, LCD lighted |
|
Probe type: |
all DSP single-axis laboratory probes |
|
Temperature range: |
5°C - 40°C |
|
Analogue outputs: |
corrected and non-corrected ± 3,5 V |
|
Interfaces: |
USB |
|
Power supply: |
100 - 240 V AC, 50 / 60 Hz |
|
Functions: |
Max Hold, Alarm with Relay, Zero Probe, Relative Reading, Auto Range, AC/DC, Keylock |
|
Accessories: |
Zero gauss chamber |
|
Special accessories: |
probe cable in lengths 3 / 7,6 / 15,24 / 30,49 m |
Standard accessories
Recommended options
GAU-005002 Zero field chamber BZC-101
GAU-005004 Additional axial probe BAP-101
GAU-005005 Additional transversal probe BTP-101
GAU-005003 Aluminum hard case
Hall probes overview Gaussmeter
Axial Probes
|
Model |
L |
D |
A |
Stem Material |
Corrected accuracy (%) of reading |
Temperature coefficient (maximum) zero |
Temperature coefficient (maximum) calibration |
|
MMA-2502-VH |
2 " ± 0.063" |
0.25" ± 0.006'' |
0.010" |
Aluminum |
± 0.25 % to 30 kG |
± 0.09 G/°C |
- 0.04 %/°C |
|
MNA-1904-VH |
4' ± 0.125" |
0.187" ± 0.005'' |
0.005'' |
Fiberglass | |||
|
MMA-2502-VG |
2" ± 0.063'' |
0.25" ± 0.006'' |
0.010" |
Aluminum |
± 0.15 % to 30 kG |
± 0.13 G/°C |
- 0.005 %/°C |
|
MNA-1904-VG |
4' ± 0.125" |
0.187" ± 0.005'' |
0.005" |
Fiberglass | |||
Transverse Probes
|
Model |
L |
D |
A |
Stem Material |
Frequency Range |
Corrected accuracy (%) of reading |
Temperature coefficient (maximum) zero |
Temperature coefficient (maximum) calibration |
|
MMT-6J04-VH |
4"± 0.125" |
0.061" max |
0.180" |
Aluminum |
DC |
± 0.25 % to 30 kG |
± 0.09 G/°C |
- 0.04 %/°C |
|
MNT-4E04-VH |
0.045" max |
0.150'' |
Rigid glass epoxy |
DC and 10 Hz to 400 Hz | ||||
|
MMT-6J04-VG |
0.061" max |
0.180" |
Aluminum |
DC |
± 0.15 % to 30 kG |
± 0.13 G/°C |
- 0.005 %/°C | |
|
MNT-4E04-VG |
0.045" max |
0.180" |
Rigid glass epoxy |
DC and 10 Hz to 400 Hz | ||||
Flexible Transverse Probes
|
Model |
W |
T |
Active Area |
Stem Material |
Frequency Range |
Corrected accuracy (%) of reading |
Temperature coefficient (maximum) zero |
Temperature coefficient (maximum) calibration |
|
MFT-3E03-VH |
0.135" max |
0.025" max |
0.040" dia (approx) |
Flexible |
DC and 10 Hz to 400 Hz |
± 0.25 % to 30 kG |
± 0.09 G/°C |
- 0.04 %/°C |
|
MFT-3E03-VG |
± 0.15 % to 30 kG |
± 0.13 G/°C |
- 0.005 %/°C |
More products BROCKHAUS
|
PRECISION LABORATORY TESLAMETER BROCKHAUS GM-475
• Precision laboratory Gaussmeter for determining the magnetic field strength of permanent magnets and systems |
|
THREE-CHANNEL PRECISION LABORATORY TESLAMETER BROCKHAUS GM-460
• Three-channel Gaussmeter for determining the magnetic field strength of permanent magnets and systems |
|
PRECISION LABORATORY TESLAMETER BROCKHAUS GM-455 DSP
• Precision laboratory Gaussmeter for determining the magnetic field strength of permanent magnets and systems |
|
GAUSSMETER TESLAMETER LABORATORY HIGH PRECISION BROCKHAUS BGM-201
• Digital precision Gaussmeter for determining the magnetic field strength of permanent magnets and systems |
|
GAUSSMETER TESLAMETER FOR MEASURE MAGNETIC FIELD WITH HALL PROBE BROCKHAUS BGM-101
• From DC to 10 kHz AC fields (TRUE RMS) |
|
FLUXMETER FOR MEASURING ALL MAGNETIC QUALITY PARAMETERS BROCKHAUS F10
• Magnetic flux |
|
DIGITAL LABORATORY GAUSSMETER BROCKHAUS GM-425
• Single-channel Gaussmeter with a resolution of 4 ¾ digits |
|
HELMOLTZ COIL FOR MAGNETIC MOMENT MEASURING BROCKHAUS HC01/HC02
• Helmholtz Coil HC01 150 mm |
|
HYSTOGRAPH FOR HYSTERESIS DIAGRAM, REMANENCE, FIELD STRENGHT, ENERY, POLARIZATION OF MAGNETIC MATERIALS BROCKHAUS HG200
• Remanence |
|
HYSTOGRAPH FOR TESTING PARAMETERS OF MAGNETIC MATERIALS BROCKHAUS HG200 AC/DC
• Remanence |