YOKOGAWA Clamp-on Tester CL155 (AC, True RMS, 2000A)

Cod: YOKOGAWA CL155


 

Brosura produs PDF, 1.05 Mb

Descriere

Features:

  • AC current measurement with current ranges up to 2000A
  • AC/DC voltage and resistance measurements
  • True RMS display
  • 54mm in diameter of measurable conductor
  • DC output function
  • Data hold function
  • Sleep function
  • Approved for conformity to safety standards EN61010-1, EN61010-2-031, EN61010-2-032 (CAT. HI 600 V, CAT. II1000 V)

 

Display (LCD, Digital Counts)

3999

AC Current Accuracy (400A Reference)

1.0+3(50/60Hz)
2.0+3(40~1kHz)

AC Current Accuracy (2000A Reference, 0~1500A)

2.0+3(50/60Hz)
3.0+3(40~1kHz)

AC Voltage Accuracy (40/400V/750V Reference)

1.0+2(50/60Hz)
1.5+3(40~1kHz)

DC Voltage Accuracy (40/400V/1000V Reference)

1.0+2

Resistance Accuracy (400/4k/40k/400kO reference)

1.5+2, Beeps at below 50± 35O (continuity check)

DC Output Accuracy (400A, 0~400mV Reference)

?.5% rdg ± 0.5mV (50/60Hz)
?.5% rdg ± 0.5mV (40~1kHz)

Method of detection

True RMS

Response Time

Approx. 2 seconds

Range Switching

Manual-range (on AC current range) / Auto-range (on AC voltage range, resistance range)

Data hold

On all range

Operating temperature and humidity

0-40°C, 85% RH or less (no condensation)

Influence of external magnetic field

1A or less at 400A/m

Influence of conductor position

± (2.0% rdg + 3dgt) or less

Safety Standard

Conforms EN 61010-1, EN61010-2-031, EN61010-2-032

Circuit voltage

1000Vrms or less

Withstanding voltage

5.55kV AC for one minute

Power Supply

R6P(SUM-3)x2 or LR6x2

Battery Life1

150 hours

Consumed current

Approx. 5mA

Sleep Function

Automatically powered down in about 10 minutes after the last switch operation

Diameter of measurable conductor

54mm at maximum

Dimensions

Approx. 105(W)x247(H)x49(D)mm

Weight (Approx.)

470g

Accessories

Clamp Adapter (Model 99025), Output Cable for Terminal Screw (Model 91019), Carrying Case (Model 93034), Output Plug (Model 98012), Test Lead (Model 98011) User's manual (IM CL120), batteries

Order Notes

Each item selected MUST be ordered as a separate line item.

Product Notes

All Clamp-on Meters include Test Leads (where applicable) and a carrying case.

 

Alte produse YOKOGAWA

YOKOGAWA AQ4280 SURSA OPTICA PORTABILA

• - SC, FC or other connector adapter
• (Removable for convenient access for cleaning)
• AQ4280 A/B/C : FC/PC, SC/PC, ST/PC
• - Rubber protector to prevent damage when dropped
• - Carrying pouch to house unit and connector adapters

 

YOKOGAWA AQ6150 SERIES OPTICAL WAVELENGTH METER

• High wavelength accuracy of ±0.3pm
• Cope with modulated light and optical filter measurement
• Simultaneous measurement of up to 1024 wavelengths
• Increase throughput with high speed measurement

 

YOKOGAWA AQ6375B LONG WAVELENGTH OPTICAL SPECTRUM ANALYZER 1200 - 2400 NM

• Long wavelength range: 1200 to 2400 nm
• Gas purging ports to reduce water vapor effects
• Built-In cut filter for high order diffracted light
• Double Speed Mode to further enhance fast measurement speed

 

YOKOGAWA AQ2180 POWERMETRU OPTICA PORTABIL

• Two versions of Power Meters to suit test application
• Versatile data storage and transfer
• Auto switching of measurement wavelength
• Power source

 

YOKOGAWA CLAMP-ON PROCESS METER CL420 (DC MA CURRENT)

• 0.2% Accuracy, 0.01mA Resolution
• Dual Display
• LED Torch Light, Backlight Display
• Analog Output Available

 

YOKOGAWA DIGITAL MULTIMETER TY520

• High Precision
• Basic accuracy 0.09%
• True RMS measurement
• Ensured Field Site Safety

 

YOKOGAWA - WT3000E ANALIZOR DE PUTERE DE PRECIZIE

• Basic power accuracy: 0.01% of reading
• Both 2 A and 30 A input elements can be installed in a single unit
• Good readability: 8.4-inch LCD and the range indicator LEDs
• Simultaneous measurement with 2 units (Up to 8 power input elements)
• Store function: 50 ms data storing interval

 

YOKOGAWA DIGITAL RESISTANCE METER 7556

• Wide operating rangeability (1 Ohm range to 100 M Ohm range)
• Full remote controlability via the standard RS232C serial, or optional GP-IB interfaces
• Software based calibration
• Printer output of measurement, and statistical analysis results

 

YOKOGAWA - DLM2000 OSCILOSCOP DIGITAL MIXT CU CANALE ANALOGICE SI DIGITALE

• Lineup includes 200 MHz, 350 MHz, 500 MHz bandwidth models
• Lightweight and compact
• Large 8.4-inch LCD display
• Long memory: Up to 250 M points (with /M3 option)
• High speed sampling: Up to 2.5 GS/s (1.25 GS/s with 4 ch)

 

YOKOGAWA MULTIMETRU DIGITAL PORTABIL TY710

• Conforms to safety standards EN61010-1 1000V CAT., 600V CAT.
• Safety shutters for preventing erroneous insertion of test leads into current measurement terminals
• Operating temperature range -20 to +55C
• Storage temperature range -40 to +70
• Closed Case Calibration