Code: Delta Ohm HD22569.2
Description
The instrument HD22569.2 is a bench top instrument for electrochemical measures: pH, conductivity, dissolved oxygen, and temperature. It is are fitted with a large backlighted LCD display.
The HD22569.2 measures pH, mV, redox potential (ORP) with pH, redox electrodes or electrodes with separate reference; conductivity, resistivity in liquids, total dissolved solids (TDS) and salinity with combined 4-ring and 2-ring conductivity/temperature probes with direct input or SICRAM module; concentration of dissolved oxygen in liquids (in mg/l) and saturation index (in %), using SICRAM combined probes of polarographic type with two or three electrodes and integrated temperature sensor.
The instruments is fitted with an input for the measurement of temperature with Pt100 or Pt1000 immersion, penetration or contact probes. The temperature probes are equipped with an automatic recognition module and factory calibration data are stored inside.
• The pH electrode calibration can be carried out on one or five points and the calibration sequence can be chosen from a list of 13 buffers Temperature compensation can be automatic or manual.
• The conductivity probe calibration can be performed automatically with automatically detected conductivity calibration solutions: 147µS/cm, 1413µS/cm, 12880µS/cm or 111800µS/cm or manually with calibration solutions having different values.
• The dissolved Oxygen probe's quick calibration function guarantees timely correctness of the performed measurements.
• Conductivity, dissolved oxygen and temperature probes fitted with SICRAM module can store factory and calibration data inside.
The instruments HD22569.2 is a datalogger, it can memorize up to 2,000 samples of data:
• pH or mV, conductivity or resistivity or TDS or salinity, concentration of dissolved oxygen and temperature:
The data can be transferred from the instrument connected to a PC via the multi-standard RS232C serial port and USB 2.0. The storing parameters can be configured using the menu.
The RS232C serial port can be used to transfer the acquired measurements to a 24 column portable printer in real time (HD40.1 or HD40.2).
The instruments equipped with HD22BT (Bluetooth) option can transfer data without any connection to a PC or printer fitted with Bluetooth input or through Bluetooth/RS232C converter.
The software DeltaLog11 allows instrument management and configuration, and data processing on PC.
The instruments have IP66 protection degree.
More products SENSECA
|
SILICON PHOTODIODE PYRANOMETER WITH GLOBAL SOLAR RADIATION MEASUREMENT FOR PHOTOVOLTAIC PANELS DELTA OHM LP SILICON-PYRA04
• LP Silicon-PYRA 04. pyranometer |
|
8-16 CHANNEL TEMPERATURE DATALOGGER DELTA OHM HD32.7 HD32.8.8 HD32.8.16
• HD32.7 - 8 INPUTS DATA LOGGER FOR pt100 SENSOR PROBES |
|
TWO-AXIS ULTRASONIC ANEMOMETER FOR WINDSPEED & WINDDIRECTION MEASUREMENT DELTA OHM HD52.3
• 2 axes ultrasonic Anemometers series HD 52.3D |
|
MULTIPARAMETRIC PH-METER-CONDUCTIVIMETER-THERMOMETER DELTA OHM HD2156.1 HD2156.2
• pHmeter-Conductivity-Thermometer measures pH, mV, conductivity, liquid resistivity, total dissolved solids, salinity and temperature. |
|
AIR QUALITY ANALYSIS MONITORING DATALOGGER CO CO2 DELTA OHM HD37AB17D HD37B17D
• data loggers able to measure and memorize simultaneously the following parameters: |
|
DATALOOGER WITH MICROCLIMATE ANALYSIS WBGT DELTA OHM HD32.2 HD32.2A
• analysis of WBGT index (Wet Bulb Glob Temperature: wet bulb temperature and globe thermometer temperature) in presence or in absence of solar radiation |
|
LABORATORY ELECTROCHEMICAL MULTIPARATEMER TESTER FOR DISSLVED OXIGEN, CONCENTRATION, SATURATION INDEX DELTA OHM HD3409.2
• Bench-top Oxymeter-Thermometer measures dissolved oxygen, |
|
ANEMOMETER-THERMOMETER DELTA OHM HD2103.1 HD2103.2
• Anemometer-Thermometer for air speed measurement which use hot-wire or vane probes, for temperature Pt100 sensor with SICRAM module. |
|
DUAL INPUT THERMOMETER DELTA OHM HD2178.1 HD2178.2
• Thermometer with two inputs, the first for Pt100 probes equipped with SICRAM module, the second input for thermocouple type K, J, T, E, N |
|
CENTESIMAL THERMOMETER DELTA OHM HD2107.1 HD2107.2 HD2127.1 HD2127.2
• Centesimal thermometer in the range +/-199.99°C, decimal outside this range |