Brand: Samar

Resistor Box in Steps High Values and Hilgh Voltage 5kV Samar CRC/4

Code: Samar CRC/4


 

Datasheet PDF, 0.99 Mb

Description

  • Decade box with spires
  • studied to simulate high resistance
  • suitable to check the insulation testers with test voltage up to 5000V d.c.
  • Ranges that can be made with the 4 contact spires supplied: 10x1 M?, 10x10M?, 10x100M?, 10x1G?.
  • In this way are selectable all values of resistance included between 1M? and 11.11G? with steps of 1 M? . Accuracy: ±1%.
  • Temperature coefficient: 100 ppm °C .
  • Max power 2W for each element.
  • Applicable max voltages : 1250V with 1M?, 2500V with 2M?, 3750V with 3M? and 5000V for values higher than 3M?.
  • In a case with removable cover.
 

More products SAMAR

PRECISION RESISTANCE STANDARD FOR METROLOGY CALIBRATION LABORATORY SAMAR RC80

• On request various values and precision
• Long term stability: +/- 10 ppm per year
• Resistances from 0.0001? to 10? are 4 terminals
• Resistances from 100? to 10M? are two electrolytic copper terminals with 2 screws, one for current and one for voltage
• Included certificate of calibration of our laboratory
• high-accuracy used as standard in laboratory.

 

STANDARD PRECISION CAPACITOR FOR CALIBRATION LABORATORIES SAMAR CC

• Values from 1nF to 1µF
• ON REQUEST DIFFERENT VALUES FROM THE ONES SHOWN
• Nominal voltage: 500Vdc
• Temperature coefficient: - (100 ± 50) x 10-6 /°C
• Test voltage: 2,000Vdc
• Insulation resistance: 500,000 MW
• Precision ± 0.1 % at 1,000 Hz, between terminals

 

HIGH VALUE AND HILGH VOLTAGE STANDARD RESISTORS SAMAR RC80/M

• On request various values and precision
• Nominal Values: 10MO..100TO
• Nominal Voltage 500V
• Included certificate of calibration of our laboratory

 

PRECISION RESISTANCE STANDARD FOR USE IN AIR SAMAR RC-E 50

• On request various values and precision
• Long term stability: ± 50 ppm per year
• Resistances from 0.0001? to 0.1? are 4 terminals, two potential and two current.
• Resistances from 1? to 10M? are wrapped in an ant-inductive way and are two terminals with 2 screws for potential and current
• Included certificate of calibration of our laboratory
• high-accuracy used as standard in laboratory.

 

RESISTANCE BOX IN STEPS VOLTAGE POTENTIAL DIVIDER SAMAR CRC/1

• Used as a potential divider or for calibration (for example insulation tester)
• Chemical resistances of high stability, with dissipation 1W
• Precision 1% dc and ac up to 1 MHz. (on request execution at 0.1% )
• Range: 0 to 500M? with terminals for the following values:
• 0.5 - 1 - 2 - 3 - 4 - 5 - 10 - 20 -30 - 40 - 50 - 100 - 200 - 300 - 400 - 500 K?
• 1 - 2 - 3 - 4 - 5 - 10 - 20 -30 - 40 - 50 - 100 - 200 - 300 - 400 - 500 M?
• Execution in reading-desk metal case.

 

STANDARD PRECISION INDUCTOR FOR CALIBRATION LABORATORIES SAMAR IC

• Values from 1mH to 1H
• Inductances of precision used as standard in laboratory
• SAMAR certificate of calibration is supplied with each standard inductor.
• Precision: ± 0.1 % at 1,000 Hz, between terminals
• ON REQUEST DIFFERENT VALUES FROM THE ONES SHOWN

 

STANDARD CAPACITOR WITH FIXED VALUE FOR CALIBRATING SAMAR CC-E50

• Values from 1nF to 1µF
• ON REQUEST DIFFERENT VALUES FROM THE ONES SHOWN
• Nominal voltage : 500Vdc
• Temperature coefficient: - (100 ± 50) x 10-6 /°C
• Test voltage: 2,000Vdc
• Insulation resistance: 500,000 MW
• Precision: ± 0.5 % at 1,000 Hz, between terminals

 

BRIDGE WHEATSTONE THOMSON MAXWELL DE SAUTY SCHERING

• DC / AC BRIDGES
• THOMSON - WHEATSTONE BRIDGE
• MAXWELL - DE SAUTY BRIDGE 50Hz
• SCHERING BRIDGE

 

RESISTIVE PRECISION DECADE BOXE SAMAR CR-E

• DECADES BOXES OF RESISTANCE with switch precision 0.1% at terminals
• Calibration precision: ± 0.02 %
• The precision shown refers to the terminals and therefore is total resistance included the resistance of zero
• Resistors are in manganine wire (temperature coefficient ± 10 ppm/°C) wrapped in ant-inductive way.
• Thanks to an opportune seasoning process the long term stability is ensured: : 10 ppm in the first year and max. 20 ppm in three years

 

RESISTANCE PRECISION DECADE BOXE SAMAR CR

• DECADES BOXES OF RESISTANCE with switch precision 0.05% at terminals
• Calibration precision: ± 0.02 %
• Thanks to an opportune seasoning process the long term stability is ensured: 10 ppm in the first year and max 20 ppm in three years.
• Switches have a resistance of contact lower than 0.001 W invariable in the time.
• certificate of calibration of our laboratory