10kV High Resistance Decade for calibrating Megohmmeters and Insulation Testers Meatest M191

Code: Meatest M191


 

Datasheet PDF, 4.67 Mb
     

Description

  • Total resistance range 10.00 k? to 1000.0 G?
  • Resistance accuracy 0.1 to 5 %
  • Test voltage range 10 000 VDC + 5%
  • Test voltage accuracy indication 0.5 % + 10 V
  • Test current indication 4 digit from 0.01 pA to 99.99 mADC
  • Programmable high resistance decade box
  • Short current verification
  • Dielectric and absorption parameters simulation
  • High voltage capacitance in three fix points 10, 50, 100 nF
  • HIPOT timer verification 

 

 

Description

M191 calibrator is based on programmable high resistance decade in summary range from 10 k? to 1 T?. The decade is designed for operation voltages up to 10 kV. In this range it offers basic accuracy 0.1 % to 5 % depending on set resistance value. 
Device is completed with additional electronic circuits allowing calibration not only resistance ranges, but also calibration of test voltage generated by UUT, short current verifying, verifying of functions of measuring dielectric parameters like polarization index (PI), dielectric absorption ratio (DAR) and polarization ratio (PR). The calibrator enables also verification of timer function of UUTs. 

The calibration can be controlled manually using front panel keypad or in remote mode using one of two types of interfaces GPIB, RS232. The calibrator can easily fit within calibration systems featuring CALIBER software support.

 

 

More products MEATEST

THREE-PHASE POWER & ENERGY CALIBRATOR MEATEST M133C

• Voltage "phase-neutral" 1 V to 280 Vdc / 300 mA, 1 V to 600 Vac
• AC voltage "phase - phase" 2 V to 1200 V
• DC/AC current 8 mA to 30 A / 5 V
• DC/AC current 90 A in one phase configuration
• Power factor setting -1 to +1
• Floating current outputs to 450 Vpk
• Harmonic, interharmonic distortion, modulation
• Frequency range 15 Hz to 1 kHz
• Simulated electric power to 54 kVA (2.7 MVA with option 140-50)
• Built-in process multimeter
• GPIB & RS232 interface

 

UNIVERSAL MULTIFUNCTION CALIBRATOR MEATEST M141

• 100ppm Accuracy
• DC/AC voltage up to 750 V
• DC/AC current up to 2 A
• Fix standard resistors from 10 O to 10 MO in decade values
• TC temperature sensors simulation
• Frequency output up to 20 MHz
• RS232 & GPIB (option) interface

 

LCR IMPEDANCE CALIBRATOR OF RESISTANCE, CAPACITANCE, INDUCTANCE MEATEST M550

• Resistance standards from 0.1 O to 100 MO
• Capacitance standards from 10 pF to 100 µF
• Simulated inductance standards from 10 µH to 10 H
• Four pair coaxial terminal connection, four terminal connection, two terminal connection
• Calibration data from 20 Hz to 1 MHz
• Interface bus GPIB and RS232 as standard

 

PRECISION LABORATORY CALIBRATOR FOR VOLTAGE, CURRENT, RESISTANCE, CAPACITY, FREQUENCY MEATEST M140

• 35ppm Accuracy
• DC/AC voltage up to 1000 V
• DC/AC current up to 20 A
• Resistance up to 50 MO, capacitance up to 50 µF
• TC/RTD temperature sensors simulation
• Frequency output up to 20 MHz
• Electric power up to 240 V / 10 A
• Built-in process multimeter
• GPIB & RS232 interface

 

THREE-PHASE POWER & ENERGY CALIBRATOR MEATEST M133CI

• Voltage "phase-neutral" 1 V to 280 Vdc / 300 mA, 1 V to 600 Vac
• AC voltage "phase - phase" 2 V to 1200 V
• DC/AC current 8 mA to 30 A / 5 V
• DC/AC current 90 A in one phase configuration
• Power factor setting -1 to +1
• Floating current outputs to 450 Vpk
• Frequency range 15 Hz to 1 kHz
• Simulated electric power to 54 kVA (2.7 MVA with option 140-50)
• Built-in process multimeter
• GPIB & RS232 interface

 

RESISTANCE CAPACITANCE INDUCTANCE STANDARD FOR CALIBRATING LCR-METERS

• Resistance Capacitance Inductance standards
• Traceability of electric quantities
• Calibration of meters
• Calibration uncertainty 0.005 %
• Low time constant
• Three/four terminal or four pair terminal connection

 

LCR IMPEDANCE CALIBRATOR OF RESISTANCE, CAPACITANCE, INDUCTANCE MEATEST M550

• Resistance standards from 0.1 O to 100 MO
• Capacitance standards from 10 pF to 100 µF
• Simulated inductance standards from 10 µH to 10 H
• Four pair coaxial terminal connection, four terminal connection, two terminal connection
• Calibration data from 20 Hz to 1 MHz
• Interface bus GPIB and RS232 as standard

 

AC/DC RESISTANCE LOAD FOR CALIBRATION OF VOLTAGE SOURCES, TRANSFORMERS MEATEST M192

• Resistance range from 15 O to 300 kO
• Maximal load 3kW, maximal voltage 250 Vrms
• Accuracy 0.1 %
• Temperature coefficient < 10 ppm/°C
• Interface RS - 232 (IEEE488 optionally)
• Table version (19" module as option)

 

PRECISION PROGRAMMABLE RESISTANCE DECADE BOX MEATEST M630

• Resistance range 1.000 00 O - 1.200 000 MO
• Accuracy 0.003%
• Maximal voltage 200Vpk
• Resolution from 10 µO
• Range of Pt sensor temperature simulation -200°C - +850°C
• Connection 2, 3, 4-wire
• Remote control RS232, optionally USB, IEEE488, Ethernet

 

PRECISION RESISTANCE DECADE BOX MEATEST M622

• Resistance range 1.000 00 O - 1.200 000 MO
• Accuracy 0.005%
• Resolution from 10 µO
• Range of Pt sensor temperature simulation -200°C - +850°C
• Connection 2, 3, 4-wire
• Remote control RS232, optionally IEEE488