  • Measures RF Current Without Direct Connection

 

Electrical Specifications 91197-1 Current Probe

Frequency Minimum

10 kHz

Frequency Maximum

8 MHz

Load Impedance

50 +/- j0

Maximum Primary Current Amps

400 Hz: 350
DC-60 Hz: 350
Pulse: 100
RF(CW) : 100

Transfer Impedance

0.33 ?

 

 

Electrical Specifications 91550-1 Current Probe

Frequency Minimum

10 kHz

Frequency Maximum

100 MHz

Load Impedance

50 +/- j0

Maximum Primary Current Amps

400 Hz: 350
DC-60 Hz: 350
Pulse: 100
RF(CW) : 42

Transfer Impedance

5 ?

 

 

Electrical Specifications 91550-2 Current Probe

Frequency Minimum

10 kHz

Frequency Maximum

150 MHz

Load Impedance

50 +/- j0

Maximum Primary Current Amps

400 Hz: 225
DC-60 Hz: 350
Pulse: 100
RF(CW) : 28

Transfer Impedance

1 ?

 

 

 

Electrical Specifications 91550-5 Current Probe

Frequency Minimum

10 kHz

Frequency Maximum

200 MHz

Load Impedance

50 +/- j0

Maximum Primary Current Amps

400 Hz: 275
DC-60 Hz: 350
Pulse: 100
RF(CW) : 2

Transfer Impedance

1 ?

Electrical Specifications 93511-1 Current Probe

Frequency Minimum

10 kHz

Frequency Maximum

30 MHz

Load Impedance

50 +/- j0

Maximum Primary Current Amps

400 Hz: 350
DC-60 Hz: 350
Pulse: 200
RF(CW) : 100

Transfer Impedance

1 ?

 

 

 

Electrical Specifications 93686-1 Current Probe

Frequency Minimum

10 kHz

Frequency Maximum

30 MHz

Load Impedance

50 +/- j0

Maximum Primary Current Amps

400 Hz: 350
DC-60 Hz: 350
Pulse: 200
RF(CW) : 200

Transfer Impedance

2 ?

 

Electrical Specifications 93686-8 Current Probe

Frequency Minimum

10 kHz

Frequency Maximum

200 MHz

Load Impedance

50 +/- j0

Maximum Primary Current Amps

400 Hz: 300
DC-60 Hz: 300
Pulse: 62
RF(CW) : 62

Transfer Impedance

8 ?

 

Electrical Specifications 94106-1 Current Probe

Frequency Minimum

100 kHz

Frequency Maximum

450 MHz

Load Impedance

50 +/- j0

Maximum Primary Current Amps

400 Hz: 200
DC-60 Hz: 200
Pulse: 50
RF(CW) : 20

Transfer Impedance

1-10 ?

 

 

Electrical Specifications 94111-1 Current Probe

Frequency Minimum

1 MHz

Frequency Maximum

1000 MHz

Load Impedance

50 +/- j0

Maximum Primary Current Amps

400 Hz: 200
DC-60 Hz: 200
Pulse: 50
RF(CW) : 20

Transfer Impedance

1-10 ?

                      

Electrical Specifications 94111-2 Current Probe

Frequency Minimum

1 MHz

Frequency Maximum

1000 MHz

Load Impedance

50 +/- j0

Maximum Primary Current Amps

400 Hz: 200
DC-60 Hz: 200
Pulse: 150
RF(CW) : 1

Transfer Impedance

1 ?

 

Electrical Specifications 94430-1 Current Probe

Frequency Minimum

10 kHz

Frequency Maximum

250 MHz

Load Impedance

50 +/- j0

Maximum Primary Current Amps

400 Hz: 200
DC-60 Hz: 200
Pulse: 70
RF(CW) : 16

Transfer Impedance

6 ?

 

 

Electrical Specifications 94606-1 Current Probe

Frequency Minimum

10 kHz

Frequency Maximum

100 MHz

Load Impedance

50 +/- j0

Maximum Primary Current Amps

400 Hz: 350
DC-60 Hz: 350
Pulse: 100
RF(CW) : 100

Transfer Impedance

5 ?

 

 

Electrical Specifications 94606-6 Current Probe

Frequency Minimum

1 kHz

Frequency Maximum

5 MHz

Load Impedance

50 +/- j0

 

More products ETS-LINDGREN

POWER FILTERS FOR RFI/ EMI RADIATION PROTECTION WITH LOW EARTH LEAKAGE CURRENTS 6..100A ETS-LINDGREN N242X

• Suitable for mobile installations
• Extremely low current leakage
• Over 100dB insertion loss from 150 kHz to 10 GHz
• Fully accessible end chambers
• RoHS compliant and CE marked for compliance with the low voltage directive

 

GTEM TEST CELL FOR EMC RADIATED EMISSIONS & IMMUNITY ETS-LINDGREN 5402

• Frequency: DC - 20 GHz (RI) 2 9 kHz - 5 GHz (RE) 1
• Post Production
• For all Phases of EMC Testing - Design Qualification - Pre-Compliance
• Full Compliance IEC 61000-4-3 ANSI C63.4

 

HORIZONTAL/ VERTICAL ELECTRIC FIELD GENERATOR 10 KHZ..30 MZ ETS-LINDGREN 5502

• Generates vertical or horizontal electric field
• Generates Fields 200 v/m in the 10 kHz to 30 MHz Frequency Range
• Fully Adjustable Height and Separation of Main Conductors
• Large Immunity Volume
• Chamber Supported

 

PASIVE LOOP SHIELDED ANTENNA FOR SHIELDING AND IMMUNITY TESTING 10 KHZ..30 MHZ ETS-LINDGREN 6512

• Frequency Minimum 20 Hz
• Frequency Maximum 5 MHz
• Connectors BNC (f)
• Maximum Input Power 20 W
• Pattern Type omnidirectional
• Polarization linear

 

ACOUSTIC TEST ENCLOSURE FOR SMALL DEVICES ETS-LINDGREN SD

• Acoustically Controlled Test Environment
• Survey and/or Engineering Grade Measurement
• Applicable Test Standards: ISO 30744, GM W7293, GM W14011

 

HIGH PERFORMANCE POWER FILTERS FOR TEMPEST/ EMI/ EMC TEST CURRENTS 16..200A ETS-LINDGREN N500X

• Low power dissipation
• Compact Rugged Construction
• 100dB insertion loss from 40kHz
• Suitable for 400Hz supplies
• Current Range 16 to 200 Amperes

 

MAGNETIC SHIELDED COIL ANTENNA LOW FREQUENCIES 20 HZ..500 KHZ ETS-LINDGREN 7604

• Frequency Minimum 20 Hz
• Frequency Maximum 500 kHz
• Connectors Type BNC
• Impedance (Nominal) 50 ?
• Pattern Type omnidirectional
• Polarization linear

 

LINEAR POLARIZED LOG PERIODIC HIGH POWER ARRAY ANTENNA 20..220 MHZ ETS-LINDGREN 3151

• 20 MHz to 220 MHz frequency range
• High power feed handles up to 5 kW RF input power
• Space saving design features tophat-loaded, low-frequency elements
• Mounted on wheeled base for easy mobility

 

MAGNETIC FIELD GENERATING COIL ANTENNA 20 HZ..50 KHZ ETS-LINDGREN 7603

• Frequency Minimum 20 Hz
• Frequency Maximum 50 kHz
• Connectors Banana Jack (2)
• Impedance (Nominal) 50 ?
• Maximum Input Power 5 A
• Pattern Type omnidirectional
• Polarization linear

 

DOUBLE-RIDGED MICROWAVE GUIDE HORN ANTENNA 750 MHZ..18 GHZ ETS-LINDGREN 3115

• Frequency Minimum 750 MHz
• Frequency Maximum 18 GHz
• Connectors Type N (f)
• Impedance (Nominal) 50 ?
• Maximum Continuous Power 500 W
• Pattern Type directional
• Polarization linear